Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-35
MEMory:SELect
Syntax:
MEMory[:TABLe]:SELect
space
<ANALOGout,VPROPF1,VPROPF2>
Example:
MEM:SEL VPROPF1
! SELECTS THE V
PROP
F MEMORY TABLE VPROPF1
! FOR EDITING IN FOLLOWING LINES
Description:
This command selects between one of three editable memory tables. This is the first line
required for configuring the V
PROP
F input. This command line must be immediately followed by
the V
PROP
F start frequency configuration, MEMory:FREQuency, then the V
PROP
F slope
configuration, MEMory:SLOPe.
Only two of the memory tables, VPROPF1 and VPROPF2, are valid selections for V
PROP
F
configuration. The ANALOGout memory table is valid only for configuration of the meters
Analog Out feature (see Section 3.2.27 for details)
Using the meter’s automated V
PROP
F capability to correct for Cal Factor requires configuring
the V
PROP
F IN connector to match the signal that source will output. This is performed by
setting the frequency so that source’s V
PROP
F output is 0.0 V. Then the slope relationship is
entered in the form of a value representing the V/Hz relationship. If, for example, source is a
Giga-tronics GT-2520B Series Signal Generator, this relationship is 0.5 V/GHz or 5e-8 V/Hz.
MEMory:FREQuency
Syntax:
MEMory[:TABLe]:FREQuency
space
<Start frequency>
Example:
MEM:FREQ 1E9
! SETS V
PROP
F FREQUENCY TO 1 GHZ
Description:
This command sets the automated V
PROP
F correction start frequency. This is the frequency at
which the voltage input into the 8650B is 0.0 V. This is the second line required for configuring
the V
PROP
F input. This command line must follow the V
PROP
F memory table configuration,
MEMory:SELect.
MEMory:SLOPe
Syntax:
MEMory[:TABLe]:SLOPe
space
<Volts per Hz>
Example:
MEM:SLOP 1E-9
! SETS V
PROP
F SLOPE TO 1 V/GHZ
Description:
This command sets the V/Hz relationship for automated Cal Factor correction. This is the final
line required for configuring the V
PROP
F input. This command line must follow the V
PROP
F start
frequency configuration, MEMory:FREQuency.
3.2.17.1
Cal Factor Error Control
•
See Error Messages in Section 3.2.36
•
Selected basic SCPI syntax and execution errors apply to these commands
•
Device-specific errors include the following and other -300 level errors
•
A common device-specific error occurs when the frequency sent to the 8650B in the
SENS#:CORR:FREQ
### command is outside the sensor operating frequency range.
For example, sending SENS1:CORR:FREQ 18.4e9 when an 18 GHz (max) 80301A CW Power Sensor is
attached will yield a device-specific error.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...