Series 8650B Universal Power Meters
2-42
Publication 31470-001, Rev C, November 2, 2017
2.4.18
Peak Power Measurements
Peak power sensors directly measure the amplitude of pulsed microwave signals. The direct sampling technique
is more accurate than traditional duty cycle correction methods. The sample position can be displayed on an
oscilloscope.
1.
Calibrate a peak power sensor and connect it to a pulsed microwave source.
2.
From the Main Menu, press
[Sensor Setup]
. From the Sensor Setup menu, select the Peak sensor with the
cursor keys.
3.
Press
[Config]
to display the Setup mode menu and select
[Peak]
.
4.
Select the desired trigger level (for internal or external triggering).
5.
Select the desired sample delay (for internal or external triggering).
6.
Optionally, set the desired delay offset (for internal or external triggering).
7.
Connect the peak power sensor’s Detector Out to an oscilloscope to view the sample position. For 80350A
Peak Power Sensors, also connect the sensor’s Sample Delay output to the oscilloscope and trigger on that
channel.
2.4.19
Measuring an Attenuator (Single Channel Method)
Attenuators are useful for many applications. With the 8650B, attenuators can be calibrated quickly and
accurately. The single channel calibration procedure outlined below is efficient for calibrating at a single
frequency or at a limited number of frequencies.
1.
Connect the power sensor to the signal source through a 6 dB attenuator (a matching pad) and adjust the
source output power to about 0 dBm. Verify that the source output is stable.
2.
Press
[FREQ]
and enter the operating frequency (this step is optional).
3.
From the Main menu, press
[Rel]
to set the reference level.
4.
Insert the attenuator to be calibrated between the matching pad and the power sensor.
5.
Record the attenuator value.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...