Series 8650B Universal Power Meters
3-32
Publication 31470-001, Rev C, November 2, 2017
CALCulate?
Syntax:
CALCulate<channel 1 to 4>[:FUNCtion]?
Example:
CALC2?
! QUERIES CHANNEL 2 SENSOR CONFIGURATION
Response:
POW #, RAT #,#, or DIF #,#
Description:
This command returns the current channel configuration. This command is a query format only.
CALCulate:UNIT
Syntax:
CALCulate<channel 1 to 4>:UNIT[:POWer]
space
<data units selection, DBM or W>
Example:
CALC1:UNIT W
! SELECTS CHANNEL 1 LINEAR UNITS, WATTS
Example 2:
CALC2:UNIT DBM
! SELECTS CHANNEL 2 LOG UNITS, DBM
Description:
This command configures a channel to report power measurements in either linear watts units
or logarithmic dBm units. When UNIT is DBM, the RATio configuration produces
measurements in dB, and POWer and DIFF measurements in dBm. When UNIT is Watts, the
RATio configuration measurements are percentages, and POWer and DIFF measurements are
in watts.
CALCulate:STATe
Syntax:
CALCulate<channel 1 to 4>:STATe
space
<ON | OFF>
Example:
CALC1:STAT OFF
! DISABLES CHANNEL 1 MEASUREMENT
Example 2:
CALC1:STAT?
! QUERIES CHANNEL 1 MEASUREMENT STATUS
Description:
This command turns a specified channel ON or OFF. For a single-channel meter, Channel 1 is
default ON. For a dual-channel meter, Channel 1 and Channel 2 are default ON. During
NORMal mode, measurement speed increases when only one sensor is performing
measurements.
Query format is available.
CALCulate:CrestFACtor:STATe
Syntax:
CALCulate<channel 1 to 4>CrestFACtor:STATe
space<
ON | OFF>
Example:
CALC1:CFAC:STAT ON
! ENABLES CHANNEL 1 CREST FACTOR MEASUREMENTS
Description:
Enables or disables Crest Factor measurements on the specified channel. Sending ’ON’ when
Crest Factor is already enabled causes the value to be reset. The Crest Factor function holds
on to the highest instantaneous power measured from the time the function is enabled until it is
reset.
CALCulate:CrestFACtor:DATa?
Syntax:
CALCulate<channel 1 to 4>CrestFACtor:DATa?
Example:
CALC3:CFAC:DAT?
! READS THE CHANNEL 3 CREST FACTOR MEASUREMENT
Description:
Reads the Crest Factor measurement from the specified display channel.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...