Introduction
Publication 31470-001, Rev C, November 2, 2017
1-11
1.3.3
Measurement Rates
The following Table illustrates typical maximum measurement rates for different measurement collection
modes. The rate of measurement depends on several factors including the controller speed and the number of
averages.
Table 1-2: Measurement Rates
Measurement Collection Mode
Readings per Second
(CW Measurement)
Readings per Second
(MAP, PAP, BAP Measurement)
Normal: Freerun (TR3)
300
150
Swift Mode
1
: Freerun or
Buffered Data w/Bus or TTL trigger
1750
N/A
Fast Buffered Mode
1
: Buffered Data,
Time Interval = 0 (fastest possible collection)
26,000
N/A
GPIB Fast Modulated Mode
1
: Freerun
N/A
800
Individual data points are read immediately after measurement in the Normal Freerun mode (continuous single
readings). The Normal Freerun mode and the GPIB Swift Freerun mode both slow down at low power levels (<-
37 dBm for Standard Sensors) to average the effects of noise. The GPIB Swift Buffered mode allows triggering of
individual data points and can store the data in the 8650B memory.
Measurement speed increases significantly using the 8650B GPIB Fast Buffered mode. Storing data in the power
meter’s memory for later downloading to the controller reduces GPIB protocol overhead. Up to 5000 readings
can be buffered.Measurement timing of individual data points with GPIB Fast Buffered mode is controlled by
setting the time interval (0 to 50 ms) between the data points following a trigger. The GPIB Fast Buffered mode
speed reflects a time interval of 0 (fastest possible collection) and does not include bus communication time.
Notes:
1.
Available only over GPIB.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...