Series 8650B Universal Power Meters
3-22
Publication 31470-001, Rev C, November 2, 2017
SENSe:CONFig:BAP
Syntax:
SENSe<sensor 1 or 2>:CONFig:BAP
Example:
SENS1:CONF:BAP
! ENABLES BAP MODE FOR SENSOR 1
Description:
Configures a modulation sensor for BAP (Burst Average Power) mode on the specified input.
This command does not affect the burst start or end exclude or dropout tolerance parameters.
The default values, set by *RST, correspond to the “automatic” settings in manual mode.
Syntax:
SENSe<sensor 1 or 2>:CONFig:BAP:BEEXclude
space
<samples>
(Where <samples> is an integer in the range of 0 to 127)
Example:
SENS1:CONF:BAP:BEEX 2
! SETS THE BURST END EXCLUDE TO 2 SAMPLES (54 µs)
Example 2:
SENS1:CONF:BAP:BEEX?
! READS THE NUMBER OF BURST END EXCLUDE SAMPLES
Description:
Sets the number of burst end exclude samples. The timing for each sample is approximately 27
μs.
Syntax:
SENSe<sensor 1 or 2>:CONFig:BAP:BSEXclude
space
<samples>
(Where <samples> is an integer in the range of 0 to 1565)
Example:
SENS1:CONF:BAP:BSEX 1
! SETS THE BURST START EXCLUDE TO 1 (27 µs)
Example 2:
SENS1:CONF:BAP:BSEX?
! READS THE NUMBER OF BURST START EXCLUDE SAMPLES
Description:
Sets the number of burst start exclude samples. The timing for each sample is approximately
27 μs.
Syntax:
SENSe<sensor 1 or 2>:CONFig:BAP:BDTolerance
space
<dropout>
(Where <dropout> is the dropout time in milliseconds in the range of 0 to 3.4 and with a resolution of 0.001 ms. The system selects
the nearest value from a series of discrete values)
Example:
SENS1:CONF:BAP:BDT 0.054
! SETS THE BURST DROPOUT TOLERANCE TO 54 µS
Example 2:
SENS1:CONF:BAP:BDT?
! READS THE CURRENT BURST DROPOUT TOLERANCE DURATION
Description:
Sets the burst dropout tolerance duration in milliseconds.
SENSe:CONFig:PAP
Syntax:
SENSe<sensor 1 or 2>:CONFig:PAP
Example:
SENS1:CONF:PAP
! ENABLES PAP MODE FOR SENSOR 1
Description:
Configures a modulation sensor for Synchronized PAP (Pulse Average Power) mode on the
specified input. This mode requires duty cycle configuration.
Syntax:
SENSe<sensor 1 or 2>:CONFig:PAP:DCYCle
space
<Duty Cycle Value>
(Where <Duty Cycle Value> is in the range 0.001 to 99.999% with a resolution of 0.001.)
Example:
SENS1:CONF:PAP:DCYC 54.54
! SETS THE DUTY CYCLE TO 54.54%
Example:
SENS1:CONF:PAP:DCYC?
! QUERIES THE DUTY CYCLE
Description:
Sets the duty cycle value that will be used when PAP mode is enabled.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...