Series 8650B Universal Power Meters
3-88
Publication 31470-001, Rev C, November 2, 2017
3.7.4
Channel Designation
The command is required to specify which channel (display line) will be assigned to all subsequent commands
that are affected.
The affected commands are: LG, LN, RE, RL(0/1/2), LH, LL, LM(0/1), MAX, MIN, MN(0/1) HIST, and
STAT(ON/OFF).
Syntax:
[AE or BE] CH
n
EN
[AE or BE] prefix specifies Sensor A or Sensor B. This is optional as some channel operations
are not specific to a sensor. One example is channel resolution, RE.
CH is the function command for Channel Designation
n
variable specifies the channel number (display line) designation where
n
= 1,2,3 or 4
EN is the required terminating suffix
Example:
BE CH 3 EN
! THE COMMANDS TO FOLLOW APPLY TO SENSOR B AND DISPLAY LINE 3
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...