Series 8650B Universal Power Meters
2-24
Publication 31470-001, Rev C, November 2, 2017
2.4.3
Sensor Calibration & Zeroing
All sensors must be calibrated and zeroed before making measurements.
2.4.3.1
Calibration & Zeroing
1.
Connect the sensor to be calibrated from Channel A or B to the Calibrator output (See Figure 2-14).
NOTE:
Connector sensor to the Calibrator connector after the power meter is turned on. Allow a 15-
minute thermal equilibrium period for the sensor and calibrator connector to minimize temperature
differences.
2.
Calibrate the sensor(s) within the 5 °C to 35 °C (41 to 95 °F) operating temperature range of the Calibrator.
Figure 2-14: Setup for Sensor Calibration
3.
Press the
[CAL/ZERO]
hardkey. The meter will automatically zero and calibrate the sensor. A sensor requires
several minutes to fully zero and calibrate. After the zero/calibration routine is complete, the display will
return to the Main menu.
NOTE:
If the calibrator is set to 0.00 dBm and turned on before or immediately after calibration, it is
possible to measure approximately ±0.02 dBm from the calibrator output instead of 0.00 dBm (0.4%
difference). This is normal and within specification due to a combination of tolerances resulting from
calibrator accuracy (±1.2% worst case for one year, over temperature range of 5 °C to 35 °C) and
connection tolerance.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...