Series 8650B Universal Power Meters
2-30
Publication 31470-001, Rev C, November 2, 2017
In BAP mode, the 8650B automatically determines which portions of the signal are in the pulse and which are
not. In computing the average power, the 8650B uses only those portions that are within the pulse. The result is
that, independent of the signal’s pulse duty cycle, the meter always reads the average power in the pulse or
burst. As with the PAP mode, when measuring a pulse modulated signal with 50% duty cycle, the reading in the
BAP mode would be 3 dB higher than in the MAP mode. However, in the BAP mode, the signal’s duty cycle can
change dynamically in time without affecting the meter reading. In the PAP mode, the duty cycle factor must be
entered to match the duty cycle of the pulsed signal.
NOTE:
BAP Mode requires a minimum pulse on or off time as determined by the power sensor pulse
width specification. See Section B.2.2, Modulated Sensor Specifications, for details.