Series 8650B Universal Power Meters
3-122
Publication 31470-001, Rev C, November 2, 2017
3.19.5.1
Enabling/Disabling an Offset
The commands which enable and disable the offset function are based on the function codes OF0 and OF1:
Syntax:
[AE or BE] [OF0 or OF1]
[AE or BE] prefix specifies Sensor A or Sensor B
OF0 deactivates the offset;
OF1 activates the offset
Example:
AE OF0
! DISABLES OFFSET FOR SENSOR A
Example 2:
BE OF1
! ENABLES OFFSET FOR SENSOR B
3.19.5.2
Setting an Offset Value
The commands which specify the offset value are based on the OS function code:
Syntax:
[AE or BE] OS [
n
] EN
[AE or BE] specifies Sensor A or Sensor B
OS indicates that an offset is being specified for the sensor
[
n
] specifies the offset in dB. The value of
n
can range from -99.999 dB to +99.999 dB
EN is a required terminating suffix
Example:
AE OS 20.00 EN
! SETS +20.00 DB OFFSET FOR SENSOR A
Example 2:
BE OS -15.12 EN
! SETS -15.12 DB OFFSET FOR SENSOR B
NOTE:
A change to the offset of a sensor will reset any Peak Hold or Crest Factor measurement involving
that sensor.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...