Series 8650B Universal Power Meters
1-12
Publication 31470-001, Rev C, November 2, 2017
1.3.4
Remote Operation
GPIB, USB, and LAN Interface
All front panel operations and some remote-only operations can be remotely programmed using SCPI
(Standard Commands for Programmable Instruments), IEEE 488.2 and IEC-625 coding.
GPIB Interrupts
SRQs are generated for the following conditions: Power Up, Front Panel key actuation, Operation
Complete and Illegal Command.
1.3.5
GPIB Fast Buffered Mode Controls
Trigger Source
TTL or GPIB
Data Buffer Control
Pre- or Post-measurement data is collected immediately either before or after receipt of the TTL or GPIB
trigger.
Time Interval
A programmed time parameter controls time interval in milliseconds between measurements. Accurate
to 5%, typical.
1.3.6
Inputs/Outputs
V
PROP
F Input (BNC)
Recalls cal factors using source V
PROP
F output. Corrects power readings for sensor frequency response
using sweeper voltage output. Input resistance = 50K. Does not operate in the fast measurement
collection modes (normal mode only).
Analog Output A & B (BNC)
Two configureable outputs that provide an output voltage of 0 to 10V from either Channel A or Channel
B in either Log or Lin units. Does not operate in the GPIB fast measurement modes.
Trigger Input (BNC)
Accepts a TTL trigger input signal for GPIB Swift Buffered and GPIB Fast Buffered measurement modes.
The TTL trigger input signal is also used for Time Gating modes.
GPIB Connector (IEEE-488)
Interfaces the power meter to a host computer for remote programming using SCPI, IEEE 488.2 and IEC-
625 coding, also emulation modes.
USB Connector (Type B)
Interfaces the power meter to a host computer for remote programming using SCPI, IEEE 488.2 and IEC-
625 coding.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...