Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-7
4. Press
[OK]
to accept the changes or
[Cancel]
to cancel the changes and return to the Power Meter
Configuration menu.
V/F In
The V/F (V
PROP
F) input accepts a frequency referenced to 0 Vdc, which the power meter uses to determine and
apply the appropriate correction factors (Stored in the sensor EEPROM). The voltage input is supplied by a
V/GHz output from the signal source. Two values must be defined for V
PROP
F: The frequency at 0 Volts, specified
in Hz, MHz or GHz, and the scale factor specified in V/GHz. The V/GHz output connector on the frequency source
is usually labeled with the scale factor.
From the Power Meter Configuration menu, move the cursor to the
V/F In
option and press
[Config]
. The V/F
Config menu will display.
1.
Frequency:
a.
Press
[Zero]
to highlight the first digit of the frequency.
b.
Press the up/down cursor keys to select the value of the first digit of the frequency.
c.
Continue in the same manner to set all of the digits for the desired frequency. The units adjust
automatically as entering the frequency.
2.
Scale:
a.
Set the scale factor by pressing
[Scale]
and setting the value in the same manner described for
the frequency.
b.
Press
[OK]
to accept the changes or
[Cancel]
to cancel the changes, and return to the Power
Meter Configuration menu.
Analog Out A & Analog Out B
The Analog A and B outputs are voltage proportional to measured power that can be applied to auxiliary test
equipment such as a data recorder.
From the Power Meter Configuration menu, move the cursor to the
Analog Out A
or
Analog Out B
option and
press
[Config]
. The Analog Out Config menu will display.
1.
Select Source
The choices of output source are Line 1 through Line 4. The configuration of the Analog outputs is
interactive with the Display Configuration. If any line is set for accumulated data, such as Min/Max,
PKHold, CrFact, or Mean & Std Dev, it will not be displayed here for configuration.
2.
Log/Lin
The mode choices are Log and Linear. Press
[Log/Lin]
to toggle between the Log and Linear modes.
3.
Set Scale
The Set Scale menu adds the options to set the High and Low range in values of volts and power. Press
[Set Scale]
to enter this menu.
a.
Press
[Set High]
to select the high range or
[Set Low]
to select the low range.
b.
Press
[Power]
or
[Volts]
to toggle between the volts and power values.
c.
Use the cursor keys to set the desired value in either the high or low range.
d.
Press
[Set Volts/Set Power]
to toggle between each value. Set the digit value of each with the
cursor keys.
e.
Press
[OK]
to accept the changes or
[Cancel]
to cancel the changes and return to the Power
Meter Configuration menu.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...