Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-107
3.16.2
Group Execute Trigger
The GPIB GET command (Group Execute Trigger) causes all the devices on the interface, which are currently
addressed to listen, to start a device dependent operation (usually a measurement). Three simple commands
(consisting of one of three function codes and valid only over the GPIB interface) regulate the 8650B response to
a GET command:
Syntax:
[GT0 or GT1 or GT2]
3.16.2.1
Group Trigger Hold (GT0)
This command disables the response of the 8650B to a GPIB GET command.
Example:
GT0
! GROUP EXECUTE TRIGGER CANCEL
3.16.2.2
Group Trigger Immediate (GT1)
This mode is similar to the mode specified by the TR1 command (trigger immediate), except that the GT1
command causes the 8650B to wait for a GPIB GET command. When the GET command is received, it triggers a
single reading which is added to the internal digital filter. An ENTER statement will return the updated filter
power level. After a GT1 command, the instrument is placed in the standby mode.
Example:
GT1
! GROUP EXECUTE TRIGGER SINGLE MEASUREMENT
3.16.2.3
Group Trigger Immediate with Full Averaging (GT2)
This mode is similar to the mode specified by the TR2 command (trigger immediate with full averaging), except
that the GT2 command causes the 8650B to wait for a GPIB GET command. When the GET command is received,
it triggers a new series of readings; enough to update the digital filter for a noise free reading at the current
power level. An ENTER statement will return the fully updated filter power level. After a GT2 command, the
instrument is placed in the standby mode.
Example:
GT2
! GROUP EXECUTE TRIGGER SINGLE MEASUREMENT WITH FULL SETTLING
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...