Series 8650B Universal Power Meters
2-22
Publication 31470-001, Rev C, November 2, 2017
Modulation Sensors
Modulation sensors can be used in CW, Modulated Average Power (MAP), Peak Average Power (PAP) and Burst
Average Power (BAP) modes.
1.
From the Sensor Setup menu select the Modulation sensor with the cursor keys.
2.
Press
[Config]
to select and configure a modulation sensor. The options for this configuration are
CW
,
MAP
,
PAP
, and
BAP
.
a.
Press
[CW]
to select the CW mode. There is no configuration for modulation sensors operating in the
CW mode and the screen will return to the Sensor Setup menu to configure other sensors.
b.
Press
[MAP]
to select the MAP mode. There is no configuration required and the screen will return to
the Sensor Setup menu to configure other sensors.
c.
Press
[PAP]
to select the PAP mode. A prompting will occur to enter a duty cycle for pulse average
inputs. Use the cursor keys to adjust the Duty Cycle. The range is 0.001% to 99.999%.
d.
Press
[BAP]
to select the BAP mode. The options are Auto Config and User Config.
1.)
Press
[Auto Config]
for the meter to automatically configure itself and return to the Sensor Setup
menu.
2.)
Press
[User Config]
to configure the meter manually. A prompting will occur to enter the values for
the
Burst Start Exclude
,
Burst End Exclude
and
Dropout
. Enter these values by pressing the
corresponding softkey and using the cursor keys to change the values (Refer to Sections 2.4.15 and
2.4.16 for details on these BAP mode parameters).
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes, and return to the Sensor Setup
menu to configure the next sensor, or return to the Main menu.
NOTE:
Operating the 8651/8652B Universal Power Meters with the 807XXA Sensors in a high level RF
Field (Approximately 3 V/m) may degrade performance. This degradation occurs at measured levels
below -30 dBm and when the frequency of the field is nominally between 100 and 200 MHz.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...