Series 8650B Universal Power Meters
3-138
Publication 31470-001, Rev C, November 2, 2017
Table 3-22: Entry Error Code Returned in Position aa
Error Code Message
Notes
00
No Error
All OK
50
Entered Cal Factor out of range
Enter value between 1.0% and 150.0%
51
Entered Offset out of range
Enter value between -99.999 dB and +99.999 dB
53
Entered average number out of range
Enter valid average number
54
Entered recall memory number out of range
Enter valid recall memory number between 0 and 20
55
Entered store memory number out of range
Enter valid store memory number between 1 and 20
56
Entered Ref cal factor out of range
Send for service
57
Memory error or battery fail
Send for service
60
Unable to set requested measurement mode or sensor
measurement because sensor A is unattached or uncalibrated
This may be due to receipt of a MAP, PAP, BAP,
DC1or DY command
61
Unable to set requested measurement mode or sensor
measurement because sensor B is unattached or uncalibrated
This may be due to receipt of a MAP, PAP, BAP,
DC1or DY command
62
Unable to set up sensor A to perform a modulated
measurement (MAP, PAP, or BAP), because the sensor is not a
modulated sensor. OR: Unable to modify a BAP measurement
because the sensor is not a modulated sensor
This may be set due to receipt of a MAP, BAP, DC1,
DY, BSTE, BSPE or BTDP command
63
Same as Error 62, but for sensor B
67
Unable to activate Peak Hold or Crest Factor features
This may be set due to receipt of a CR1 or PH1
command. Peak Hold and Crest Factor can be
enabled only in MAP, PAP or BAP modes
68
Unable to initiate fast measurement collection mode
Verify using CW sensor and measurement mode
selections
69
Unable to perform two-sensor operation (AR, BR, AD, BD)
Verify two sensors are attached and calibrated
70
Entered peak sensor A data error
Check entered data
71
Entered peak sensor B data error
Check entered data
72
Entered peak sensor A delay out of range
Check entered delay
73
Entered peak sensor B delay out of range
Check entered delay
74
Entered peak sensor A trigger out of range
Check entered trigger value
75
Entered peak sensor B trigger out of range
Check entered trigger value
76
Sensor EEPROM data entry has error
Check entry data
77
Sensor A does not exist
Check sensor A. This error code refers only to the
EEPROM command
78
Sensor B does not exist
Check sensor B. This error code refers only to the
EEPROM command
79
Measurement settling target for auto-average mode is out of
range
Value must be 0.10% to 100.00%
81
Duty cycle out of range
Value must be between 0.001% and 99.999%
82
Frequency value out of range
Value must be between 0 Hz and 100 GHz
85
Resolution value out of range
Value must be between 0 and 3
90
GPIB data parameter error
Check, then enter with valid parameter
91
Invalid GPIB code
Check, then enter with correct code
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...