Inspecting with the Sensitive Search Inspection Item
270
FQ2-S/CH User’s Manual
Measurement Data That Can Be Logged
The following values can be logged as measurement data.
*
When logging data is output, the data is output in the order of the above table.
7-6 Logging Measurement Data and Image Data
Errors
Errors in Teaching
A teaching error message will appear if the contrast of the image within the model registration region is too low.
Select a region with a larger contrast between light and dark areas compared to the region that was registered
as the model and re-register it as the model.
Parameter
Range of value
Description
Judgement
−
2: No judgment (not measured)
0: Judgment is OK
−
1: Judgment is NG
This is the judgment result of the sensitive search.
Correlation
0 to 100
Correlation
D deviation
Color image: 0 to 221
Monochrome image: 0 to 127
Density deviation
Position X
−
99,999.9999 to 99,999.9999
This is the X coordinate of the position where the
model was found.
Position Y
−
99,999.9999 to 99,999.9999
This is the Y coordinate of the position where the
model was found.
Angle
−
180 to 180
This is the angle of the position where the model was
found.
NG sub-region
0 to 100
This is the number of NG sub-regions.
Sub-region number
0 to 99
This is the number of the regions with the lowest cor-
relation.
Sub-region pos. X
(sub-region number (X))
0 to 9
This is the column number in the X direction of the
output region.
Sub-region pos. Y
(sub-region number (Y))
0 to 9
This is the row number in the Y direction of the output
region.
Sub-region position X
−
999,999,999.9999 to
999,999,999.9999
This is the X coordinate of the sub-region that was
found.
Sub-region position Y
−
999,999,999.9999 to
999,999,999.9999
This is the Y coordinate of the sub-region that was
found.
Correlation
(sub-region)
0 to 100
Correlation
Deviation
(sub-region)
Color image: 0 to 221
Monochrome image: 0 to 127
Density deviation
Summary of Contents for FQ2-S/CH Series
Page 1: ...User s Manual Smart Camera FQ2 S CH Series Cat No Z337 E1 06...
Page 12: ...10 FQ2 S CH User s Manual MEMO...
Page 74: ...Setting Up Ethernet 72 FQ2 S CH User s Manual MEMO...
Page 366: ...Editing the Model Region Measurement Region from Run Mode 364 FQ2 S CH User s Manual MEMO...
Page 432: ...Functions Related to the System 430 FQ2 S CH User s Manual MEMO...
Page 587: ...Index FQ2 S CH User s Manual 585 Index 9...
Page 589: ......