6 List Sequence Measurements
Save
Key Path
Save
Remote Command
:MMEMory:STORe:CORRection 1|2|3|4|5|6, <filename>
Example
:MMEM:STOR:CORR 2 "myAmpcor.csv" saves Correction 2 to the file myAmpcor.csv on the current
path.
The default path is My Documents\amplitudeCorrections.
Notes
If the save is initiated via SCPI, and the file already exists, the file will be overwritten.
Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.
Both single and double quotes are supported for any filename parameter over SCPI.
Dependencies
Corrections are not supported by all Measurements. If in a Mode in which some Measurements
support it, this key will be grayed out in measurements that do not. The key will not show at all if no
measurements in the Mode support it.
This key will not appear unless you have the proper option installed in your instrument.
Readback
Selected Correction
Backwards
Compatibility SCPI
:MMEMory:STORe:CORRection ANTenna | CABLe | OTHer | USER, <filename>
For backwards compatibility, ANTenna maps to 1, CABle maps to 2, OTHer maps to 3 and USER
maps to 4
Initial S/W Revision
A.02.00
Correction Data File
A Corrections Data File contains a copy of one of the analyzer correction tables. Corrections provide a way
to adjust the trace display for predetermined gain curves (such as for cable loss).
Corrections files are text files in .csv (comma separated values) form, to make them importable into Excel
or other spreadsheet programs. The format for Corrections files is as follows.
Line
#
Type of field
Example
Notes
1
File type, must be “Amplitude
Correction”
Amplitude Correction
May not be omitted
2
File Description (in quotes)
“Correction Factors for 11966E”
60 characters max; may be empty but
may not be omitted. If exceeds 60
characters, error –233 Too much data
reported
3
Comment (in quotes)
“Class B Radiated”
60 characters max; may be empty but
may not be omitted. . If exceeds 60
characters, error –233 Too much data
reported
4
Instrument Version, Model #
A.02.06,N9020A
May be empty but may not be omitted
5
Option List, File Format
Version
K03 LFE EXM ,01
May be empty but may not be omitted
6
Freq Unit to be used for all
Frequency Unit,MHz
assumed to be Hz if omitted
Sequence Analyzer User's & Programmer's Reference
505
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...