6 List Sequence Measurements
Source
When you use a power reference, the signal generator outputs an RF power that is set relative to the
reference power by the value entered under Source>Amplitude>RF Power as follows:
Output power = reference power + entered power
Where:
reference power equals the original RF Power entered under Source>Amplitude>RF Power and set as the
reference power
entered power equals a new value entered under Source>Amplitude>Amptd Offset
For more information on Reference Frequency refer to
"Set Reference Power " on page 515
Key Path
Source, Amplitude
Remote Command
:SOURce:POWer:REFerence <ampl>
:SOURce:POWer:REFerence?
:SOURce:POWer:REFerence:STATe OFF|ON|0|1
:SOURce:POWer:REFerence:STATe?
Example
:SOUR:POW:REF 0.00 dBm
:SOUR:POW:REF:STATe ON
Dependencies
This setting is unavailable and is grayed out when the
is turned ON.
Couplings
This value is coupled to the
"Set Reference Power " on page 515
key such that pressing the Set
Reference Power key updates the reference power with the current output power.
Preset
0.00 dBm
OFF
Min
–125.00 dBm
Max
10.00 dBm
Initial S/W Revision
A.05.00
Amptd Offset
Allows you to specify the RF output power offset value.
When the amplitude offset is set to zero (0) and you set a new offset value (positive or negative), the
displayed amplitude value will change as follows and the RF output power will not change:
Displayed value = output power + offset value
Where:
output power equals the original RF Power entered under Source>Amplitude>RF Power
offset value equals the value entered under Source>Amplitude>Amptd Offset
When the amplitude offset is set to a value other than zero (0) and you enter a new RF power value under
Source>Amplitude>RF Power, the displayed power will be the same as the value entered and the RF
output power will be equal to the value entered minus the offset value as follows:
Output power = entered power – offset power
516
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...