6 List Sequence Measurements
Meas Setup
Basic IQ Data Setup
This allows you to return a limited amount of IQ data samples back as measurement results.
Averaging is not allowed when this measurement is selected in an Acquisition. If averaging is set, a
"settings conflict" error is returned.
Basic IQ Data has two measurement parameters: "Digital IF BW" and "MeasType", which apply to the
entire Sequence.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35179
Digital IF BW
Activates the Digital IF BW function, which enables you to manually set the information bandwidth of the
analyzer. This is used to set the digital IF filter of the EXT Analyzer.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup, Basic IQ Data Setup
Remote
Command
[:SENSe]:LSEQuencer:BIQData:DIF:BANDwidth[:RESolution] <freq>
[:SENSe]:LSEQuencer:BIQData:DIF:BANDwidth[:RESolution]?
Example
LSEQ:BIQD:DIF:BAND 25 MHz
LSEQ:BIQD:DIF:BAND?
Couplings
If different Digital IF BW were set for several measurements in an acquisition (HW State) then the List Sequencer
SW would have to choose the highest value and grab the acquisition. Basic IQ data returns the captured samples
WITHOUT re-sampling.
Preset
25 MHz
State Saved
Saved in instrument state.
Min
1 kHz
Max
40 MHz
Help Map
ID
35194
Measurement Type
This setting is used to set the Basic IQ measurement to return raw IQ samples or Amplitude/Phase pairs.
Sequence Analyzer User's & Programmer's Reference
461
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...