6 List Sequence Measurements
Meas Setup
between a previous acquisition completing and the trigger occurring). This value applies to all non-Free
Run triggers in a sequence. When the STATE is set to ON and the trigger arming period is less than the trig
timeout value, then the measurement continues and the trig timeout is activated on the next non-Free Run
trigger in the sequence. No timeout error is thrown for that trigger event.
If the trigger arming period is greater than the trig timeout value then a trigger timeout error is thrown, and
the integrity value for that acquisition is returned as “trigger error”. The measurement is only aborted if
“Abort on Error” is enabled. Otherwise the measurement continues to wait for a trigger, and if a trigger
event occurs, the measurement proceeds and trigger timeout is activated on the next non-Free Run trigger
in the sequence.
When the STATE is set to OFF, then no trigger timeout error is thrown, and this parameter is disabled.
parameter_table_23.52941
52.94118
Key Path
Meas Setup
Remote Command
[:SENSe]:LSEQuencer:TIMeout:TRIGger <time>
[:SENSe]:LSEQuencer:TIMeout:TRIGger?
[:SENSe]:LSEQuencer:TIMeout:TRIGger:STATe OFF|ON| 0|1
[:SENSe]:LSEQuencer:TIMeout:TRIGger:STATe?
Example
LSEQ:TIM:TRIG 500ms
Couplings
The measurement is only aborted if “Abort on Error” is enabled.
Coupled to the integrity indicator “Trigger Error”.
Preset
2 s
Off
State Saved
Saved in instrument state.
Range
1 us to 4 ks
Min
1 us
Max
4 ks
Initial S/W Revision
A.05.00
Help Map ID
35197
Include Source in Sequence
This parameter determines if the internal source is is incorporated into the analyzer sequence. When the
source sequence is incorporated in this way, the source sequence is initiated automatically by the analyzer
sequence, and cannot be initiated independently.
parameter_table_23.52941
52.94118
Key Path
Meas Setup
Remote Command
[:SENSe]:LSEQuencer:INCLude:SOURce YES|NO|0|1
[:SENSe]:LSEQuencer:INCLude:SOURce?
Example
LSEQ:INCL:SOUR YES
464
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...