6 List Sequence Measurements
Source
External 2
Sets the trigger input for the current step to External 2.
Key Path
Source, List Sequencer, List Sequencer Setup, Step Trigger
Example
:SOUR:LIST:STEP2:SET:INP:TRIG EXT2
Notes
SCPI is supported after A.09.40
Notes
Note: When on E6640A, trigger 2 is a bi-directional trigger port. So when trigger 2 has been
configured as OUTPUT type, choosing External 2 as the input trigger for the current step will
generate error.
Initial S/W Revision
A.05.00
Transition Time
Allows you to specify the transition time for the current step.
The transition time is the amount of time allowed for the source to settle at the current frequency or
amplitude value.
Transition Time should not be taken as additional time before or inside the Step Duration. You can set a
value for the settling time to allow the source output frequency or amplitude to become stable. Make sure
that during this period of time, you do not use the source output signal.
The following table lists recommended values for appropriate settling times to allow for changes within the
source.
Value Changed
Recommended Transition Time
Frequency
500 µs
Amplitude
100 µs to within 0.1 dB
20 µs to within 1.0 dB
If the Transition Time value is shorter than the time necessary for the hardware to settle and a List
Sequence is initiated, a warning is generated.
If the Transition Time value is longer than the Step Duration, an error is generated when initiating a source
list sequence. For source list sequence, transition time is included in the step duration length. If the
Transition Time value is longer than the Step Duration Time, the real step duration length is extended to
equal the transition time and cause a timing shift.
Key Path
Source, List Sequencer, List Sequencer Setup
Remote Command
:SOURce:LIST:STEP[1]|2|3…1000:SETup:TRANsition:TIME <time>
:SOURce:LIST:STEP[1]|2|3…1000:SETup:TRANsition:TIME?
Example
:SOUR:LIST:STEP2:SET:TRAN:TIME 1ms
:SOUR:LIST:STEP2:SET:TRAN:TIME?
Sequence Analyzer User's & Programmer's Reference
601
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...