6 List Sequence Measurements
List Sequence Measurements
This scpi command is to query if the results of the specified acquisition are available.
If it returns 0, that means the results are still in calculation. Otherwise, it means the specified acquisition
results are ready to query.
:FETCh:LSEQuencer:ASYNc:ACQuire{1:512}?/qonly/
Return the results of specified acquisition once the results are available no matter if the whole sequence is
completed or not.
This scpi command will be blocked until the results are ready.
The results definition is the same as :FETCh:LSEQuencer:ACQuire{1:512}?/qonly/
:FETCh:LSEQuencer:ASYNc:ACQuire{1:512}:ASTep{1:1000}?/qonly/
Return the results of specified analysis step once the results are available no matter if the whole sequence
is completed or not.
This scpi command will be blocked until the results are ready.
The results definition is the same as :FETCh:LSEQuencer:ACQuire{1:512}:ASTep{1:1000}?/qonly/
:INITiated:LSEQuencer:ARMed?
This scpi command provides the capability to indicate if the sequence has finished its initialization. It won’t
return until the instrument is ready to perform capture.
The MEASure query MEAS:LSEQ1|2|3? presets the test set to thedefault state and returns default results.
This has limited use since a preset returns the Sequence settings and meas setup parameters to their
default settings, and hence the results are not useful.
parameter_table_23.52941
52.94118
Key Path
Meas
Mode
Sequence Analyzer
Initial S/W Revision
A.05.00
Help Map ID
35012
Example for two Acquisitions, each containing two Analysis Steps
non_
parameter_
table_
8.63562
7.794118
65.9232
Condition
N
Results Returned
Mode =
Sequence
Analyzer
Not
specifiedor
n=1
For example, 2 Acquisitions with 2 Analysis Steps in each, the first step in Acq1 making a GSMK
PFER measurement and the second step in Acq1 making a Basic Transmit Power and WCDMA Mod
Acc, Acq2 returns the same as Acq1 but at a different freq. The results would return the following:
1 value: Total number of real values returned for the Sequence, including this one.
1 value: Reserved for Keysight use.
228
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...