6 List Sequence Measurements
Meas Setup
Example
:LSEQ:ASRL:MSNR 10
Notes
Not available for E6607C,E6640A and E6650A.
Couplings
When “Auto Set RF Levels” parameter is set to Off, this parameter is disabled.
Preset
15
State Saved
Saved in instrument state.
Range
0 dB to 70 dB
Min
0 dB
Max
70 dB
Initial S/W Revision
A.05.00
Help Map ID
35178
Peak Power Margin
This parameter is fed into the auto set RF levels algorithm. It specifies a difference between expected peak
power and full scale, after the ranging is done. This margin is intended to accommodate the output power
uncertainty of the DUT. A default value of 5 dB is used. That means if the Auto RF level ranging is finished
successfully, you won’t get an overload warning if the actual power is less than (the Peak Power s 5
dB). Increasing the margin may cause the test set to increase its attenuation.
If the Auto Set RF Level is set to OFF, then these additional parameters are disabled.
parameter_table_23.52941
52.94118
Key Path
Meas setup, Auto Set RF Levels
Remote Command
[:SENSe]:LSEQuencer:ASRLevels:PPMargin <rel_ampl>
[:SENSe]:LSEQuencer:ASRLevels:PPMargin?
Example
:LSEQ:ASRL:PPM 5
Couplings
When “Auto Set RF Levels” parameter is set to Off, this parameter is disabled.
Preset
5
State Saved
Saved in instrument state.
Range
0 dB to 10 dB
Min
0 dB
Max
10 dB
Initial S/W Revision
A.05.00
Help Map ID
35261
Trigger Timeout
This parameter sets the timeout for trigger events that occur throughout a sequence. It can effectively be
used as a measurement timeout. If all acquisition triggers are set to Free Run, then this parameter is not
active. It should be set to a value greater than the trigger arming period (that is, the time expected
Sequence Analyzer User's & Programmer's Reference
463
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...