6 List Sequence Measurements
Source
Displayed Power = output power + offset power
Displayed power = entered power
Where:
entered power equals the amplitude entered under Source>Amplitude>RF Power
offset power equals the value previously entered and set under Source>Amplitude>Amptd Offset
Key Path
Source, Amplitude
Remote Command
:SOURce:POWer[:LEVel][:IMMediate]:OFFSet <rel_ampl>
:SOURce:POWer[:LEVel][:IMMediate]:OFFSet?
Example
:SOUR:POW:OFFS 0.00 dB
Dependencies
This setting is unavailable, and is grayed out when the List Sequencer is turned ON.
Preset
0.00 dB
Min
–200.00 dB
Max
200.00 dB
Initial S/W Revision
A.05.00
Modulation
Allows you to toggle the state of the modulation.
Key Path
Source
Remote Command
:OUTPut:MODulation[:STATe] ON|OFF|1|0
:OUTPut:MODulation[:STATe]?
Example
:OUTP:MOD OFF
Notes
This setting is for independent mode and has no effect on
. If the
is set to ON, the list sequencer controls the source output and this key will
be grayed-out. And this setting will be none-forceful grey out on front panel to indicate out-of-scope.
Non-forceful means user still can change this setting by SCPI but cannot change manually on front
panel. When setto Off will make source leave list sequencer and this setting will be black out and
take effect immediately When the Modulation is ON, the “MOD” annunciator is displayed in the
system settings panel. When the Modulation is turned Off, the “MOD” annunciator is cleared. If the
is set to ON, the “MOD” annunciator will be replaced by “SEQ” in the
system settings panel indicating that the output is controlled by list sequencer.
Preset
Off
Range
On | Off
Initial S/W Revision
A.05.00
Frequency
Allows you to access the Frequency sub-menu.
Sequence Analyzer User's & Programmer's Reference
517
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...