5 System Functions
System
For front panel operation, you are prompted to confirm action before setting the alignment parameters to
factory defaults:
The parameters affected are:
Parameter
Setting
Timebase DAC
Calibrated
Timebase DAC setting
Calibrated value
Auto Align State
Normal (if the instrument is not operating with default
alignment data, Off otherwise)
Auto Align All but RF
Off
Auto Align Alert
Time & Temperature
Key Path
System, Alignments
Mode
All
Example
:SYST:DEF ALIG
Notes
Alignment processing that results as the transition to Auto Alignment Normal will be executed
sequentially; thus *OPC? or *WAI will wait until the alignment processing is complete.
Initial S/W Revision
Prior to A.02.00
Execute Expired Alignments (Remote Command Only)
Alignments can be expired in the situation where Auto Align is in the state of Partial or Off. This feature runs
the alignments that have expired. This is different than performing an Align All, Now operation. Align All,
Now performs an alignment of all subsystems regardless of whether they are needed or not, with Execute
Expired Alignments, only the individual subsystems that have become due are aligned.
Mode
All
Remote Command
:CALibration:EXPired?
Example
:CAL:EXP?
Notes
:CALibration:EXPired? returns 0 if successful
Sequence Analyzer User's & Programmer's Reference
185
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...