6 List Sequence Measurements
Meas Setup
The settings defined in the subsection of
"Basic Transmit Power Setup" on page 451"Basic Transmit Power
take effect only when the radio standard of the acquisition (please refer to
Standard" on page 394 "Radio Standard" on page 394
) is set to None.
For selections other than None, the sample rate, channel filter shape and channel filter parameters are
fixed to provide a measurement equivalent to the GSM mode’s Transmit Power measurement, as well as
the Channel Power measurement in WCDMA/HSPA, CDMA2000, 1xEVDO, LTE-FDD, LTE-TDD ,
TDSCDMA modes:
non_parameter_table_
17.85131
13.79085
15.26144
13.95425
16.4951
Radio Standard
Digital IF BW(Hz)
Channel Filter BW
(Hz)
Channel Filter
Shape
Channel Filter
Parameter
WCDMA/HSPA
7.68 *10e6 / 1.25
3.84*10e6
RRC
Alpha = 0.22
CDMA2000
2.4576*10e6
/1.25
1.23*10e6
Short Nyquist
Alpha = 0.1
1xEVDO
2.4576*10e6 /
1.25
1.23*10e6
Short Nyquist
Alpha = 0.1
GSM/EDGE
510*10e3
510*10e3
Flattop
LTE(Both FDD and TDD)1
Bandwidth * 1.6
Bandwidth
Short Nyquist
Alpha = 0.1
TDSCDMA
1.5e6
1.5e6
RRC
Alpha = 0.22
Note 1: The bandwidth of LTE could be changeable. The current bandwidth can be derived from Mode
Setup
à
Demod
à
Bandwidth in LTE mode.
Upper Limit
This parameter sets the threshold level for the Basic Tx Power pass/fail result. When set to ON, the
measurement checks each Analysis Step relative to the “Exp Power” level set for each Analysis Step. If it is
above the “Exp Power” level plus this upper limit value, then the Pass/Fail result is “1”. If below, the
Pass/Fail result is “0”. When set to OFF, the measurement populates the Pass/Fail result with “–1”,
signifying it is untested.
For the “Input Overload”, integrity value is returned whether the Upper Limit is set to ON or OFF.
This parameter is also coupled with amplitude correction if it is enabled.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup, Basic Transmit Power Setup
Remote Command
[:SENSe]:LSEQuencer:BTXPower:LIMit:UPPer <rel_ampl>
[:SENSe]:LSEQuencer:BTXPower:LIMit:UPPer?
[:SENSe]:LSEQuencer:BTXPower:LIMit:UPPer:STATe OFF|ON|0|1
[:SENSe]:LSEQuencer:BTXPower:LIMit:UPPer:STATe?
452
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...