6 List Sequence Measurements
List Sequence Measurements
Meas Bitmap Header is a floating point number which appears ahead of the results for the analysis
interval; it is a decimal representation of the meas bitmap value for the measurements included in the
interval. The value returned for WCDMA Modulation Accuracy results is 2048. If more than one
measurement is included in this analysis step, then the value returned represents the combined meas
bitmap value for all included measurements, and the results are returned in ascending order of their meas
bitmap values.
Number Of Results is a floating point number which appears ahead of the results of a measurement within
the analysis interval; it represents the number of results metrics returned by the measurement. The value
returned for WCDMA Modulation Accuracy results is 22.
WCDMA Modulation Accuracy (Rho) measurement returns the following comma-separated scalar results:
non_
parameter_
table_
15.52288
51.47059
Index
Result Parameter
0
Overall Pass/Fail Result shows the mask test result
0:Pass, 1:Fail, –1:Not tested
1
RMS EVM pass/fail result (1.0 = fail, 0.0 = pass)
2
Peak EVM pass/fail result (1.0 = fail, 0.0 = pass)
3
Frequency Error pass/fail result (1.0 = fail, 0.0 = pass)
4
Rho pass/fail result (1.0 = fail, 0.0 = pass)
5
Peak Code Domain Error pass/fail result (1.0 = fail, 0.0 = pass)
6
RMS EVM (Average) is a floating point number (in percent) of EVM over the entire measurement area
7
Peak EVM (Peak Hold) is a floating point number (in percent) of the Peak EVM in the entire measurement area.
8
Magnitude Error (Average) is a floating point number (in percent) of the average magnitude error over the entire
measurement area.
9
Phase Error (Average) is a floating point number (in degree) of the average phase error over the entire
measurement area.
10
I/Q Origin Offset (Average) is a floating point number (in dB) of the I and Q error (magnitude squared) offset from
the origin
11
Frequency Error (Average) is a floating point number (in Hz) of the frequency error in the measured signal
12
Rho (Average) is a floating point number of Rho
13
Peak Code Domain Error (Peak Hold) is a floating point number (in dB) of the Peak Code Domain Error relative to
the mean reference power
14
Peak Code Domain Error Channel Number (Peak Hold) is the channel number in which the peak code domain
error is detected
15
Number of active channels
16
Time Offset (Average) is a floating point number (in chips) of the pilot phase timing from the acquisition trigger
point
17
Total power over a slot (Average) is a floating point number (in dBm) of the total RF power over a measurement
slot
Sequence Analyzer User's & Programmer's Reference
251
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...