5 System Functions
System
not put into the SCPI error queue by default. Normally the status bit is the only way for detecting these
errors remotely.
It is possible to enable Input Overload reporting to the SCPI queue, by issuing the
:SYSTem:ERRor:OVERload ON command. To return to the default state, issue the
:SYSTem:ERRor:OVERload OFF command. In either case, Input Overloads always set the status bit.
For versions of firmware before A.10.01, the Input Overload was only a Warning and so was never available
in the SCPI queue, although it did set the status bit. For A.10.01 and later, the Input Overload is an error
and can be enabled to the SCPI queue using this command.
Key Path
SCPI only
Remote Command
:SYSTem:ERRor:OVERload[:STATe] 0|1|OFF|ON
Example
:SYST:ERR:OVER 1 Enable overload errors
Preset
Set to OFF by Restore Misc Defaults (no Overload errors go to SCPI)
State Saved
Saved in instrument state.
Initial S/W Revision
A.10.01
Power Up (Remote Command Only)
This serves to show the errors encountered during the application boot-up, such as: mismatch FW-FPGA,
missing Calibration data, missing hardware and construction errors.
Remote Command
:SYSTem:ERRor:PUP?
Notes
If no error occurs, the return value will be: “No Power Up Errors.”
Return Value: <list of error strings>.
<List of error strings> is an <IEEE488 Block> format.
Return Value Example:
“Power up errors, see details in Windows Event Log”
"Unmatched FPGA Version(s), See details in Windows Event Log”
Initial S/W Revision
E.14.30
System
The System screen is formatted into three groupings: product descriptive information, options tied to the
hardware, and software products:
156
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...