6 List Sequence Measurements
Source
options for specifying these are:
<filename> - plays the specified waveform from the start. The filename value is the name of the file
within ARB playback memory, it is does not include the windows path to the file on the HDD. If you
enter a filename for a waveform that does not reside within ARB playback memory, an error is
generated.
CONT – continues playback of the ARB file from the previous step
CW – outputs a CW tone
OFF – disable RF output
8. Step Duration <
enum
> – specifies the duration of the step. The duration can be specified to be
either time, or play count of the ARB file associated with the step, or continuous. If Waveform is set
to “CW”, this value cannot be set to Play Count and an error will be generated. If continuous is
selected, the following Time or Count value is ignored. For further details of this setting, see
9. Time or Count <time/int> – specifies time duration in seconds or play count of the ARB file
associated with the step. For further details of this setting, including the valid ranges for the time or
play count setting,
and
10. Output Trigger <Boolean> – specifies the output trigger for the step. For details of the ranges for
this setting see
Dependencies
The range of subopcode values is 1 to 1000 and the value you enter is determined by the number of
steps you have configured. For details see
.
If you attempt to remotely set or query a subopcode that is out of range, an error is generated.
Initial S/W Revision
A.05.00
Step Configuration of Step Trigger parameter list (Remote Command Only)
This SCPI command is to configure “Step Trigger” parameter array of the whole List Sequencer at one
time. The number of array is same as step number defined in
Number of
Steps. As a step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:INPut:TRIGger <enum>, <enum>, <enum>, …
:SOURce:LIST:SETup:INPut:TRIGger?
Example
:SOUR:LIST:SET:INP:TRIG IMM,INT,EXT2
:SOUR:LIST:SET:INP:TRIG?
Notes
The command is to setup below parameter array of whole list sequence.
Step Trigger <
enum
> – specifies the input trigger for the step. For details of the valid types of step
trigger see
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
number of steps will be updated.
Remote Command Notes
IMMediate|INTernal|KEY|BUS|EXTernal2
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see
Initial S/W Revision
A.09.40
Sequence Analyzer User's & Programmer's Reference
631
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...