6 List Sequence Measurements
Marker
marker to make this attribute relevant. If it is a Delta marker, the reference marker determines how the
marker is controlled and how its value is displayed. A marker cannot be relative to itself.
parameter_
table_
23.52941
52.94118
Key Path
Marker, Properties
Remote
Command
:CALCulate:LSEQuencer:MARKer[1]|2|...|12:REFerence <integer>
:CALCulate:LSEQuencer:MARKer[1]|2|...|12:REFerence?
Example
CALC:LSEQ:MARK1:REF 2
Notes
A marker cannot be relative to itself so that choice is unavailable, and if sent from SCPI generates error –221:
"Settings conflict; marker cannot be relative to itself."
You must be in the SEQAN mode to use this command. Use INSTrument:SELect to set the mode.
When queried a single value is returned (the specified marker numbers relative marker).
Couplings
The act of specifying the selected marker’s reference marker makes the selected marker a Delta marker.
Preset
The preset default “Relative To” marker (reference marker) is the next higher numbered marker (current marker
+1). For example, if marker 2 is selected, then its default reference marker is marker 3. The exception is marker
12, which has a default reference of marker 1.
Set to the defaults by using Restore Mode Defaults. This is not reset by Marker Off, All Markers Off, or Preset.
State Saved
Saved in instrument state. Not affected by Marker Off and hence not affected by Preset or power cycle.
Min
1
Max
12
Initial S/W
Revision
A.05.00
Help Map
ID
35251
Marker Trace
Assigns the specified marker to the designated trace.
parameter_
table_
23.52941
52.94118
Key Path
Marker, Properties
Remote
Command
:CALCulate:LSEQuencer:MARKer[1]|2|...|12:TRACe RFENvelope | MAXRfenvelop |
MINRfenvelop
:CALCulate:LSEQuencer:MARKer[1]|2|...|12:TRACe?
Example
CALC:LSEQ:MARK1:TRAC RFEN
CALC:LSEQ:MARK1:TRAC?
Notes
You must be in the SEQAN mode to use this command. Use INSTrument:SELect to set the mode.
Couplings
Max Hold RF Envelop is only available when Max Hold Trace is set to On. Min Hold RF Envelop is only available
Sequence Analyzer User's & Programmer's Reference
219
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...