2 Programming the Test Set
Common Commands
Common Commands
"Standard Event Status Enable " on page 88
"Standard Event Status Register Query " on page 88
"Identification Query " on page 89
"Operation Complete " on page 89
"Query Instrument Options " on page 90
"Recall Instrument State " on page 90
"*RST (Remote Command Only)" on page 91
"Save Instrument State " on page 91
"Service Request Enable " on page 92
"Status Byte Query " on page 92
"Wait-to-Continue " on page 93
All (Daily use)
Immediately executes an alignment of all subsystems which includes both the source and the analyzer in
the TRX module. The “All” alignment is sufficient to maintain specified performance, provided that (1) the
TRX’s internal temperature has not drifted more than +/–5 degree C since the previous alignment, and (2)
no more than 8 hours have elapsed since the previous “All” alignment., and (3) no more than 1 week has
elapsed since these three alignments have all been run: IF, RF, and Source, and (4) a 45 minute warm-up
period between power-up of the TRX and invoking the “All” alignment. The instrument stops any
measurement currently underway, performs the alignment, then restarts the measurement from the
beginning (similar to pressing the Restart key).
If an interfering user signal is present at the RF Input, the alignment is performed on all subsystems except
the RF. After completion, the Error Condition message “Align skipped: 50 MHz interference” or “Align
skipped: 4.8 GHz interference” is generated. In addition the Error Condition message “Align Now, RF
required” is generated, and bits 11 and 12 are set in the Status Questionable Calibration register.
The query form of the remote commands (:CALibration[:ALL]? or *CAL?) invokes the alignment of all
subsystems and returns a success or failure value. An interfering user signal is not grounds for failure; if the
alignment was able to succeed on all portions but unable to align the RF because of an interfering signal,
the resultant will be the success value.
Successful completion of Align Now, All will clear the “Align Now, All required” Error Condition, and clear
bit 14 in the Status Questionable Calibration register. It will also begin the elapsed time counter for Last
Align Now, All Time, and capture the Last Align Now, All Temperature.
Sequence Analyzer User's & Programmer's Reference
85
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...