6 List Sequence Measurements
Meas Setup
Max
100 dBm
Initial S/W
Revision
A.05.00
Help Map ID
35119
Input Trigger Delay
This parameter allows you to set the trigger delay for the current Acquisition.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup
Remote
Command
-To set Input Trigger Delay individually:
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:TRIGger[:INPut]:DELay <time>
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:TRIGger[:INPut]:DELay?
-To set the Intput Trigger Delay as an array:
[:SENSe]:LSEQuencer:LIST:SETup:TRIGger[:INPut]:DELay <time>,<time>,<time>,….
[:SENSe]:LSEQuencer:LIST:SETup:TRIGger[:INPut]:DELay?
The array form of the command is used to make a different setting for each acquisition in the sequence (the
number of array values must match the number of acquisitions; otherwise a "settings conflict" error will occur).
Example
:LSEQ:ACQ2:SET:TRIG:INP:DEL –200us
:LSEQ:ACQ2:SET:TRIG:INP:DEL?
:LSEQ:LIST:SET:TRIG:INP:DEL 0, –200us, –100us
:LSEQ:LIST:SET:TRIG:INP:DEL?
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
Positive values act as a trigger delay. A negative number acts as a pre-trigger. For example, if set to –200 us, the
meas offset would start 200 us prior to the trigger event. Meas interval starts after meas offset ends, which in this
case would be 200 us prior to the trigger event plus the Meas Offset value.
GMSK ORFS switching and GMSK PVT (for example) need to see the full rising edge and all the
overshoot/undershoot. A warning is displayed (see
"Validation and Setup dependencies" on page 378
) if selected
measurement in the first analysis step requires a rising edge then acquisition may need a pre-trigger to catch the
rising edge.
Preset
0 s
State Saved
Saved in instrument state.
Range
–150 ms to 500 ms
430
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...