5 System Functions
System
HiSLIP Server
Turns the HiSLIP server capability On or Off, enabling you to limit SCPI access over LAN through the HiSLIP
server.
HiSLIP stands for High Speed LAN Instrument Protocol and is part of the IVI–6.1 specification.
Here is an example of a VISA connection string used to connect to the HiSLIP Server on an X-Series
Spectrum Analyzer:
TCPIP0::a-n9030a–93016::hislip0::INSTR
In the example above, hislip0 is the HiSLIP device name that VISA users must include in their HiSLIP VISA
Address strings. Your HiSLIP device name may be different depending on your VISA settings.
Key Path
System, I/O Config, SCPI LAN
Mode
All
Remote Command
:SYSTem:COMMunicate:LAN:SCPI:HISLip:ENABle OFF|ON|0|1
:SYSTem:COMMunicate:LAN:SCPI:HISLip:ENABle?
Example
:SYST:COMM:LAN:SCPI:HISL:ENAB OFF
Preset
This is unaffected by Preset, but is set to ON with a “Restore System Defaults->Misc”
State Saved
No
Range
On | Off
Initial S/W Revision
A.11.00
SCPI Socket Control Port (Remote Command Only)
Returns the TCP/IP port number of the control socket associated with the SCPI socket session. This query
enables you to obtain the unique port number to open when a device clear is to be sent to the instrument.
Every time a connection is made to the SCPI socket, the instrument creates a peer control socket. The port
number for this socket is random. The user must use this command to obtain the port number of the control
socket. To force a device clear on this socket, open the port and send the string “DCL ” to the instrument.
If this SCPI command is sent to a non SCPI Socket interface, then 0 is returned.
Mode
All
Remote Command
:SYSTem:COMMunicate:LAN:SCPI:SOCKet:CONTrol?
Example
:SYST:COMM:LAN:SCPI:SOCK:CONT?
Preset
This is unaffected by Preset or “Restore System Defaults->Misc”.
State Saved
No
Range
0 to 65534
Min
0
Max
65534
Initial S/W Revision
Prior to A.02.00
188
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...