6 List Sequence Measurements
Source
see
Initial S/W Revision
A.09.40
Step Configuration of Duration Time or Play Count parameter list (Remote Command Only)
This SCPI command is to configure “Duration Time” or “Play Count” parameter array of the whole List
Sequencer at one time. The number of array is same as step number defined in
Number of Steps. As a step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:TOCount <time/int>, <time/int>, <time/int>, …
:SOURce:LIST:SETup:TOCount?
Example
:SOUR:LIST:SET:TOC 1s,2s,3s
:SOUR:LIST:SET:TOC?
:SOUR:LIST:SET:TOC 5,6,7
:SOUR:LIST:SET:TOC?
Notes
The command is to setup below parameter array of whole list sequence.
Time or Count <time/int> – specifies time duration in seconds or play count of the ARB file
associated with the step. For further details of this setting, including the valid ranges for the time or
play count setting,
and
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
number of steps will be updated.
If current
is “Continuous”, then generate error –221,"Settings
conflict;Cannot accept time or count input when step duration type is Continuous on step #"
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see
Initial S/W Revision
A.09.40
Step Configuration of Output Trigger parameter list (Remote Command Only)
This SCPI command is to configure “Output Trigger”parameter array of the whole List Sequencer at one
time. The number of array is same as step number defined in
Number of
Steps. As a step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:OUTPut:TRIGger <bool>, <bool>, <bool>, …
:SOURce:LIST:SETup:OUTPut:TRIGger ?
Example
:SOUR:LIST:SET:OUTP:TRIG ON,OFF,ON
:SOUR:LIST:SET:OUTP:TRIG?
Notes
The command is to setup below parameter array of whole list sequence.
Output Trigger <Boolean> – specifies the output trigger for the step. For details of the ranges for this
setting see
.
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
Sequence Analyzer User's & Programmer's Reference
637
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...