6 List Sequence Measurements
Meas Setup
The array form of the command is used to make a different setting for each acquisition in the sequence (the
number of array values must match the number of acquisitions; otherwise a "settings conflict" error will occur).
Example
:LSEQ:ACQ2:AST3:SET:TIME:INT 10ms
:LSEQ:ACQ2:AST3:SET:TIME:INT?
:LSEQ:ACQ2:LIST:SET:TIME:INT 500us, 1ms, 2ms
:LSEQ:ACQ2:LIST:SET:TIME:INT?
:LSEQ:NUMB:ACQ 2
:LSEQ:ACQ1:NUMB:AST 5
:LSEQ:ACQ2:NUMB:AST 5
:LSEQ:LIST:SET:TIME:INT 500us, 600us, 700us
:LSEQ:LIST:SET:TIME:INT? => 5e–4, 6e–4, 7e–4, 5e–4, 6e–4, 7e–4, 5e–4, 6e–4, 7e–4, 5e–4
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
Preset
500 us
State Saved
Saved in instrument state.
Range
0 to 1 s
Min
0
Max
1 s
Initial S/W
Revision
A.05.00
Help Map ID
35134
Measurement Bitmap
Allows you to select which measurements are included in an analysis step. Measurement bitmap choice
differs for each radio standard. See the table below for bitmapping. The measurements available to be
selected vary, depending on the licenses installed (for example, if you do not have the U9071A GSM_EDGE
measurement license, then the GSM and EDGE measurements are unavailable). The measurement bitmap
determines what results are returned via SCPI and in the Result Metrics View.
non_parameter_table_
23.75817
5.718954
32.78595
Measurement
Bit
Radio Standards Supported
Basic Transmit Power
0
ALL: BASIC, GSM, EDGE, WCDMA, CDMA2000, 1xEV-DO, LTE-FDD,
LTE-TDD, TDSCDMA, WLAN
Basic Frequency and Phase
Error
1
ALL: BASIC, GSM, EDGE, WCDMA, CDMA2000, 1xEV-DO, LTE-FDD, LTE-TDD,
TDSCDMA, WLAN
Basic Discrete PAvT
2
ALL: BASIC, GSM, EDGE, WCDMA, CDMA2000, 1xEV-DO, LTE-FDD, LTE-TDD,
TDSCDMA,WLAN
440
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...