6 List Sequence Measurements
List Sequence Measurements
interval. The value returned for PVT results is 16. If more than one measurement is included in this analysis
step, then the value returned represents the combined meas bitmap value for all included measurements,
and the results are returned in ascending order of their meas bitmap values.
Number Of Results is a floating point number which appears ahead of the results of a measurement within
the analysis interval; it represents the number of results metrics returned by the measurement. The value
returned for PVT results is 15.
non_
parameter_
table_
11.30719
66.92811
Index
Result Details
0
Overall Pass/Fail result shows the limit check result of Off power before, average power over –33rd ~ –14th chip,
and Off power after to their upper limit: –65 dBm, –50 dBm and –65 dBm respectively. Only when all these 3
powers are below their upper limits, can this overall pass/fail result be pass. This is the only pass fail result.
0:Pass, 1:Fail, –1:Not tested
1
Mean power is the mean power (in dBm) of the range specified by Analysis Time Slot and Measurement Time Slots
in the most recently acquired data. If averaging is on and average number is N, mean power is calculated after N
PVT traces are merged.
2
Burst width is the width of the first set of continuous active slots in the range specified by Analysis Time slot and
Measured Time Slots. If averaging is on and average number is N, burst width result is calculated by averaging N
burst width results.
3
Burst start time is the start time of the burst referring to the start of the analysis interval. If averaging is on and
average number is N, the final burst width results is calculated by averaging N burst start time results.
4
Rump up time is the time difference between 10% and 90% voltage points (relative to peak) on the positive slope of
the burst, here burst has the same meaning in Burst width. If averaging is on and average number is N, rump up
time is calculated after N PVT traces are merged.
5
Ramp down time is the time difference between 90% and 10% voltage points (relative to peak) on the negative
slope of the burst, here burst has the same meaning in Burst width. If averaging is on and average number is N,
rump down time is calculated after N PVT traces are merged.
6
Off power is calculated as the averaged power between the –48th and –16th chip before the burst start time. If
average is on and average number is N, the off power is calculated after N PVT traces are merged.
7
Maximum power is the maximum peak level in the range specified by Analysis Time Slot and Measured Time Slots
(in dBm). If average is on and average number is N, the maximum power is calculated after N PVT traces are
merged.
8
Minimum power is the minimum peak level in the range specified by Analysis Time Slot and Measured Time Slots
(in dBm). If average is on and average number is N, the minimum power is calculated after N PVT traces are
merged.
9
Trace sample interval is a floating point number representing the time between samples of the merged envelop
trace data, which could be read by using FETCh:LSEQuencer<l>:ACQuire<m>:ASTep<n>:TDPVt:TRACe?
If average is on and average number is N, this result is calculated after N PVT traces are merged.
10
Trace sample number is the number of data points in the merged envelop trace data, which could be read by using
FETCh:LSEQuencer<l>:ACQuire<m>:ASTep<n>:TDPVt:TRACe?
Only up to 2000 envelop trace points can be returned. So, the envelop trace returned may be decimated. If average
is on and average number is N, this result is calculated after N PVT traces are merged.
244
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...