6 List Sequence Measurements
List Sequence Measurements
off additional commands on GPIB until the acquisition is complete.
READ Commands:
:READ:<measurement>[n]?
•
Does not preset the measurement to the factory default settings. For example, if you have previously initiated the ACP
measurement and you send READ:ACP? it will initiate a new measurement using the same instrument settings.
•
Initiates the measurement and puts valid data into the output buffer. If a measurement other than the current one is
specified, the instrument will switch to that measurement before it initiates the measurement and returns results.
•
For example, suppose you have previously initiated the ACP measurement, but now you are running the channel
power measurement. Then you send READ:ACP? It will change from channel power back to ACP and, using the
previous ACP settings, will initiate the measurement and return results.
•
Blocks other SCPI communication, waiting until the measurement is complete before returning the results
•
If the optional [n] value is not included, or is set to 1, the scalar measurement results will be returned. If the [n] value
is set to a value other than 1, the selected trace data results will be returned. See each command for details of what
types of scalar results or trace data results are available. The binary data formats should be used when handling large
blocks of data since they are smaller and faster than the ASCII format. (FORMat:DATA)
Initial S/W Revision
Prior to A.02.00
Current Measurement Query (Remote Command Only)
This command returns the name of the measurement that is currently running.
Remote Command
:CONFigure?
Example
CONF?
Initial S/W Revision
Prior to A.02.00
Limit Test Current Results (Remote Command Only)
Queries the status of the current measurement limit testing. It returns a 0 if the measured results pass
when compared with the current limits. It returns a 1 if the measured results fail any limit tests.
Remote Command
:CALCulate:CLIMits:FAIL?
Example
CALC:CLIM:FAIL? queries the current measurement to see if it fails the defined limits.
Returns a 0 or 1: 0 it passes, 1 it fails.
Initial S/W Revision
Prior to A.02.00
Data Query (Remote Command Only)
Returns the designated measurement data for the currently selected measurement and subopcode.
272
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...