BRIEF INSTRUCTIONS 

CLEANING AND DISINFECTION OF 

DELTASTREAM HC

Xenios_HC_EN_003_2018_02 

 

Valid for: 

         ME DPHC 0001    deltastream HC, 230V  

         ME DPHC 0002    deltastream HC, 115V  

Safety instructions: 

Attention 

This system is a closed circuit. The water line should not 

be disconnected in the operating area. Do not perform 

water changes and disinfection in the operating area.

 

Carry out water changes and disinfection only in hygien-

ically clean surroundings.

 

Pay 

attention to hygienically clean work processes. (Dis-

posable gloves, protective mask, hood).

Make sure you and your staff are aware of mycobacteria 

and their effects.

Regularly check the system for microbiological contami-

nation. We recommend to test the water quality once a

month, nevertheless the system is in use or stored. 

Disconnect the mains plug before starting to clean and

disinfect the unit.

 

For all surfaces and parts use only disinfectants based on 

aldehydes, ammonium components or alcohols that do

not attack ABS plastic, PVC or PU. As far as possible, do

not use disinfectants based on phenol derivatives be-

cause they shorten the life of plastics. 

 A list of suitable 

disinfectants can be found on Page 2 of these Brief 
Instructions as well as on Page 26 of the IFU

.

Make sure no liquid enters the unit (e.g. through the ven-

tilation slots).

 

Only switch on the device if neither the surface is wetted 

nor if liquid is inside the device outside the tank. 

Only use sterile water for disinfecting the device! 

Fill sterile water into the tank only with auxiliaries that 

have been previously disinfected.

Surface cleaning and disinfection

 

Ventilat

 

ion slots 

Check the ventilation slots on the sides of the unit for dirt 
regularly, but at least every six months. Do not open the 
device yourself.

Dust deposits in the unit will reduce the efficiency of the 

system.  

Have any inside dirt removed by an authorized 

service technician. 

Surface 

 

After each water change, carry out a surface disinfection 

(device housing, hose set and connectors). 

If the device is contaminated by contact with body flu-

ids, disinfect the surface and, if necessary, the water-

carrying parts. 

Make sure that after surface disinfection no residues of 

the cleaning agent are left on the surface and the hous-

ing resp.  interior of the device is dry before use. 

For disinfecting the surface of the device, please, note 
the list of approved surface disinfectants on Page 2 of 
this Quick Guide or Page 26 of the IFU. When using the 
disinfectant, observe the manufacturer's instructions

.

Also clean and disinfect the screw thread of the filler 

neck.

Do not switch the unit on again until the disinfectant has 

evaporated fully. 

For more information, please see the deltastream HC

IFU in chapter 7 from page 25 and following.

CAUTION  

These cleaning instruction only apply in connec-
tion with the currently valid instruction for use of 
the above mentioned devices and does not re-
place it! 

Read the safety instructions 
below before starting to clean 
the unit:

 

During the water change or the 

disinfection, no oxygenator may 

be connected to the water circuit.

Pay attention to hygienic handling 
when cleaning the device! 