6 List Sequence Measurements
Source
:SOURce:LIST:SETup:WAVeform?
Example
:SOUR:LIST:SET:WAV “CW”,”Off”,”CONT”
:SOUR:LIST:SET:WAV?
Notes
The command is to setup below parameter array of whole list sequence.
Waveform <
string
> – specifies the waveform for playback during the step. The step can output either
a new ARB waveform, continue playback of the previous waveform, or output a CW tone. The options
for specifying these are:
<filename> - plays the specified waveform from the start. The filename value is the name of the file
within ARB playback memory, it is does not include the windows path to the file on the HDD. If you
enter a filename for a waveform that does not reside within ARB playback memory, an error is
generated.
CONT – continues playback of the ARB file from the previous step
CW – outputs a CW tone
OFF – disable RF output
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
number of steps will be updated.
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see
Range
”filename” | “CW” | “Off” | “CONT”
Initial S/W Revision
A.09.40
Step Configuration of Step Duration parameter list (Remote Command Only)
This SCPI command is to configure “Step Duration” parameter array of the whole List Sequencer at one
time. The number of array is same as step number defined in
Number of
Steps. As a step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:DURation:TYPE <enum>, <enum>, <enum>, …
:SOURce:LIST:SETup:DURation:TYPE?
Example
:SOUR:LIST:SET:DUR:TYPE COUN,TIME,CONT
:SOUR:LIST:SET:DUR:TYPE?
Notes
The command is to setup below parameter array of whole list sequence.
Step Duration <
enum
> – specifies the duration of the step. The duration can be specified to be either
time, or play count of the ARB file associated with the step, or continuous. If Waveform is set to
“CW”, this value cannot be set to Play Count and an error will be generated. If continuous is selected,
the following Time or Count value is ignored. For further details of this setting, see
.
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
number of steps will be updated.
Remote Command Notes
TIME | COUNt | CONTinuous
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
636
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...