4 Mode Functions
Mode
1. Close and restart the analyzer program without changing your configured preloads. This may free up
enough memory to load the requested Mode, depending on your configured preloads
2. Clear out all preloads and close and restart the analyzer program with only the requested application
preloaded, and with that application running. This choice is guaranteed to allow you to run the requested
application; but you will lose your previously configured preloads. In addition, there may be little or no
room for other applications, depending on the size of the requested application.
3. Bring up the Configure Applications utility in order to reconfigure the preloaded apps to make room for
the applications you want to run (this will then require restarting the analyzer program with your new
configuration). This is the recommended choice because it gives you full flexibility to select exactly what
you want.
4. Exit the dialog box without doing anything, which means you will be unable to load the application you
requested.
In each case except 4, this will cause the analyzer software to close, and you will lose all unsaved traces
and results.
If you attempt to load a mode via SCPI that will exceed memory capacity, the Mode does not load and an
error message is returned:
–225,"Out of memory;Insufficient resources to load Mode (mode name)"
where “mode name” is the SCPI parameter for the Mode in question, for example, SA for Spectrum
Analyzer Mode.
Sequence Analyzer
Selects the Sequence Analyzer mode for sequenced measurements. Depending on licensed applications
there may be a number of different measurements available in this mode. These measurements are all
done on IQ captured data and can be set up to calculated on any part of the capture.
If you are using the Help feature, this mode must be currently active to access its detailed information. If it
is not active, exit the Help feature (Esc key), select the mode, and re-access Help.
Key Path
Mode
Example
INST:SEL SEQAN
INST:NSEL 400
Initial S/W Revision
A.05.01
W-CDMA with HSPA+
Selects the W-CDMA with HSPA+ mode for general purpose measurements of signals following this
standard. There are several measurements available in this mode.
If you are using the Help feature, this mode must be currently active to access its detailed information. If it
is not active, exit the Help feature (Esc key), select the mode, and re-access Help.
Key Path
Mode
Example
INST:SEL WCDMA
132
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...