6 List Sequence Measurements
List Sequence Measurements
parameter_table_23.52941
52.94118
Key Path
Meas
Help Map ID
0
Integrity Indicator
The integrity indicator values are returned with the measurement results as indicated in
. The values returned are shown in the table below. The values returned are
referenced to the Acquisition or Analysis Step. Errors outside these (for example, uncal and hardware
errors that may occur outside of the Sequence) are returned over SCPI in the normal way. They appear in
the error queue and can be returned via SCPI with SYSTem:ERRor?
The integrity indicator values are coupled to the “Abort on Error” parameter if that parameter is On. In that
case the Sequence aborts if any of the integrity values are set to 1, indicating a questionable value.
The integrity indicator values are only returned via SCPI. The front panel user interface does not return the
integrity values.
non_parameter_table_
6.32353
16.19281
12.9085
42.81046
Bit
Error
Description
Level
This maps to Status Sub system bit
0
Trigger Error
Acq. Level
STATus:QUEStionable:INTegrity Trigger Error bit 11
1
Input overload
Acq. Level
STATus:QUEStionable:INTegrity IF/ADC Over Range bit 4
2
Mem Error
Acq. Level
STATus:QUEStionable:INTegrity Memory Error bit 9
3
Acq Failure
Acq. Level
STATus:QUEStionable:INTegrity Meas Error bit 8
4
Reserved
N/A
N/A
5
Reserved
N/A
N/A
6
Burst Not Found
Meas
Level
STATus:QUEStionable:INTegrity:SIGNal Burst Not Found bit
2
7
Sync Error
Meas
Level
STATus:QUEStionable:INTegrity:SIGNal Sync Error bit 6
8
Signal too noisy
Meas
Level
STATus:QUEStionable:INTegrity:SIGNal Signal too noisy bit
8
9
Insufficient Data
Meas
Level
STATus:QUEStionable:INTegrity Insufficient Data bit 7
10
Meas Error
Meas
Level
STATus:QUEStionable:INTegrity Meas Error bit 8
11
Reserved
N/A
N/A
12
Reserved
N/A
N/A
268
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...