6 List Sequence Measurements
List Sequence Measurements
11.30719
Index
Result Details
0
Pass/Fail Result shows the mask test result
0:Pass, 1:Fail, –1:Not tested
1
Returns a list of comma-separated values for the modulation spectrum at all the offsets (lower and upper.). This is
followed by the switching transient results at all the offsets (lower and upper). The carrier is considered offset zero
(0) and is the first set of results sent. Four values are provided for each of the offsets (including the carrier), in this
order:
Negative offset(a) - power relative to carrier (dB)
Negative offset(a) - absolute average power (dBm)
Positive offset(a) - power relative to carrier (dB)
Positive offset(a) - absolute average power (dBm)
Values for all possible offsets are sent. Zeros are sent for offsets that have not been defined. The total number of
values sent (120) = (4 results/offset) * * types: modulation & switching)
Carrier – modulation measurement values
Offset 1 – modulation measurement values and so on
~
Offset 14 – modulation measurement values
Carrier – switching transients measurement values
Offset 1 – switching transients measurement values
~
Offset 14 – switching transients measurement values and so on
2
Returns a list of relative level to the test limit, the relative test limit and the absolute test limit for all the offset
frequencies. The relative level to the test limit is returned for both lower and upper offsets. Four values are returned
for each offset in the following order:
1. Relative level to the test limit (dB) at the negative offset frequency
2. Relative level to the test limit (dB) at the positive offset frequency
3. Relative test limit used (dB)
4. Absolute test limit used (dBm)
Values for all possible offsets are returned.
The carrier frequency is considered offset zero (0.0 Hz) and is the first set of values returned.
Zeros are returned for offsets that have not been defined.
Zeros are returned for the measurement that was not performed. For example, if Meas Type is Modulation, all
switching transient measurement results are 0.0.
The total number of values returned is:
120 = (4 results / offset) * (15 offset frequencies) * (2 measurement types)
Carrier (Offset A) – modulation measurement results
Offset 1 (Offset B) - modulation measurement results
. . . . . .
Offset 14 (Offset O) - modulation measurement results
Carrier (Offset A)– switching transients measurement results
Offset 1 (Offset B) – switching transients measurement results
240
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...