6 List Sequence Measurements
List Sequence Measurements
Key Path
SCPI only
Remote
Command
[:SENSe]:LSEQuencer:ACQuire[1]|2|...|4..512:ASTep[1]|2|...|4…1000:SETup
<time>, <time>, <integer>, <ampl>
Example
:LSEQ:ACQ1:ASTEP1:SETup 20,20,1, –10
:LSEQ:ACQ1:ASTEP1:SETup?
Notes
The parameters are:
<
time
> - specifies the Analysis Offset for the Analysis Step.
<
time
> - specifies the Analysis Interval for the Analysis Step.
<
integer
> - specifies the Measurement Bitmap for the Analysis Step (this is the decimal equivalent of the
bits selected).
<
ampl
> - specifies the Expected Power at DUT Output for the Analysis Step.
Dependencies
The range of sub op code values is determined by the Number of Acquistions and the Number of Analysis
Steps you have configured.
If you attempt to remotely set or query a sub op code combination that is out of range, this results in an
error message.
Initial S/W
Revision
A.05.00
Help Map ID
35232
Validation and Setup dependencies
When a Sequence is transferred into the test set using a TSV file an initial verification phase is carried out.
If the file is found to be invalid, an error “–232, Invalid format, syntax error on line 0” is generated.
For all other setup methods, the validation checks prior to run time are listed below:
Validation Errors:
1. Range checking for all parameters.
2. Settings Conflict: Channel Number is not valid for the selected Band and link direction.
3. Settings Conflict: Radio Standard and Bitmap do not match (for example, a selected measurement is not
available for the selected standard).
4. Step numbers are not sequential (for example, the numbers cannot be 1,2,3,4,7).
5. Settings Conflict: Acquisition duration is not long enough to accommodate the analysis step offset and
analysis step interval.
6. Settings Conflict: Radio band not licensed.
7. For averaging, expected power must be the same for each “blue bar” (analysis interval).
8. Selected radio band not applicable to the selected radio standard.
9. Acquisition overlap pre-trigger error (acquisitions cannot overlap in time).
10. Analysis step number exceeded, due to the number of averages.
378
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...