6 List Sequence Measurements
Meas Setup
changes). This time applies to the acquisition after the Acquisition Duration.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup
Remote
Command
-To set Trainsition Time individually:
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:TIME:TRANsition <time>
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:TIME:TRANsition?
-To set the Transition Time as an array:
[:SENSe]:LSEQuencer:LIST:SETup:TIME:TRANsition <time>,<time>,<time>,….
[:SENSe]:LSEQuencer:LIST:SETup:TIME:TRANsition?
The array form of the command is used to make a different setting for each acquisition in the sequence (the
number of array values must match the number of acquisitions; otherwise a "settings conflict" error will occur).
Example
:LSEQ:ACQ2:SET:TIME:TRAN 10ms
:LSEQ:ACQ2:SET:TIME:TRAN?
:LSEQ:LIST:SET:TIME:TRAN 500us,1ms,2ms
:LSEQ:LIST:SET:TIME:TRAN?
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
The times set here are dependent on what Acquisition parameters are changing between acquisitions. For
accurate measurements the following setup rules should apply.
Instrument Gain Changes : 20us
Frequency changes:
Within same Frequency Band switching: 300us
Different Frequency Band switching: 1ms
Frequency Band 1: –0.08GHz to 0.6075GHz
Frequency Band 2: 0.5075GHz to 2.1775GHz
Frequency Band 3: 2.0775GHz to 3.6GHz
Preset
20 us
State Saved
Saved in instrument state.
Range
1 us to 4 ks
Min
0 s
Max
4 ks
Initial S/W
Revision
A.05.00
Help Map ID
35111
Sequence Analyzer User's & Programmer's Reference
425
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...