6 List Sequence Measurements
26.
..repeat 5 to 25 for Acq 2 and for all subsequent Acq’s in the Sequence
The format of the results for each measurement is detailed in the following
sections. Each measurement returns all results when queried remotely.
If the Sequence is aborted manually, then NANs are returned.If the Sequence is
aborted due to “Abort on Limit Fail”, then results return NANs from point of
failure and valid results prior to that. To find out where Limit failed use
FETC:LSEQ3?
If the Sequence is aborted due to “Abort on
Error”, then results return NANs from point
of failure and valid results prior to that. To
find out where error occurred use
FETC:LSEQ3?
Mode =
Sequence
Analyzer
2
Returns a pass fail < AcqNum>|0 for the entire Sequence. This returns a fail if any
meas limit fails OR if an error occurs. “0” means entire Sequence passed limits
and no error occurred. If fail then the number returned is the Acquisition number
(AcqNum) of first acquisition to fail a limit OR first acquisition that invoked an
error.
If the Sequence is aborted manually, then
results return –1.
Mode =
Sequence
Analyzer
3
Returns three integer numbers, Acquisition number, followed by Analysis Interval
Number, followed by the Meas Bitmap Number of first detected position to fail a
limit OR that invoked an error. Note: since the measurements are made in
multiple threads the first limit fail or error to be detected may not necessarily be
the first one sequentially.
If the Sequence is aborted manually, then
results return –1, –1, –1.
For an example of the results returned see
"Example for two Acquisitions, each containing two Analysis
e .
In addition an Acquisition’s results can be queried on its own using:
:FETCh:LSEQuencer[1]:ACQuire[1]|2|3|4..512?
:READ:LSEQuencer[1]:ACQuire[1]|2|3|4..512?
:MEASure:LSEQuencer[1]:ACQuire[1]|2|3|4..512?
This returns all result metrics for the specific acquisition “AcqNum”. Performing a READ initiates the full
Sequence and returns only the AcqNum’s results. If the Sequence is aborted manually, then results return
NANs. If the Sequence is aborted due to “Abort on Limit Fail”, then results return NANs from point of failure
and valid results prior to that. If Sequence is aborted due to “Abort on Error”, then results return NANs from
210
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...