6 List Sequence Measurements
Meas Setup
Peak Power
This parameter sets the peak power for the acquisition which is used for the Auto Set RF Level algorithm
when Auto Set RF Level is set to ON. (Note: Auto Set RF Level is always ON for E6607C, E6630A, and
E6640A). If OFF this peak power parameter is ignored.
parameter_
table_
23.57026
52.94934
Key Path
Meas Setup, Acquisition Setup
Remote
Command
-To set Peak Power individually:
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:PPOWer <amp>
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:PPOWer?
-To set Peak Power as an array:
[:SENSe]:LSEQuencer:LIST:SETup:PPOWer <amp>,<amp>,<amp>,….
[:SENSe]:LSEQuencer:LIST:SETup:PPOWer?
The array form of the command is used to make a different setting for each acquisition in the sequence (the
number of array values must match the number of acquisitions; otherwise a "settings conflict" error will occur).
Example
:LSEQ:ACQ2:SET:PPOW 20 dBm
:LSEQ:ACQ2:SET:PPOW?
:LSEQ:LIST:SET:PPOW 0,10dBm,15dBm
:LSEQ:LIST:SET: PPOW?
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
Couplings
Coupled to Auto Set RF Level when set to On. If Off this parameter is ignored.
Preset
0 dBm
State Saved
Saved in instrument state.
Min
–100 dBm
Max
100 dBm
Initial S/W
Revision
A.05.00
Help Map ID
35109
Transition Time
This parameter sets the amount of time that is required for any changes in Acquisition Parameters that
occur between Acquisitions (this allows time for transition between Frequency and or Power range
424
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...