6 List Sequence Measurements
List Sequence Measurements
There is only one value returned which represents the actually detected slot number of LTE Modulation
Accuracy.
parameter_table_23.52941
52.94118
Key Path
Meas
Initial S/W Revision
A.09.40
Help Map ID
0
TD-SCDMA Modulation Accuracy (Rho) Results
TD-SCDMA Modulation Accuracy (Rho) Results are detailed here.
The SCPI-only header values are as follows:
Meas Bitmap Header is a floating point number which appears ahead of the results for the analysis
interval; it is a decimal representation of the meas bitmap value for the measurements included in the
interval. The value returned for TD-SCDMA Modulation Accuracy results is 2048. If more than one
measurement is included in this analysis step, then the value returned represents the combined meas
bitmap value for all included measurements, and the results are returned in ascending order of their meas
bitmap values.
Number Of Results is a floating point number which appears ahead of the results of a measurement within
the analysis interval; it represents the number of results metrics returned by the measurement. The value
returned for TD-SCDMA Modulation Accuracy results is 25.
TD-SCDMA Modulation Accuracy (Rho) measurement returns the following comma-separated scalar
results:
non_
parameter_
table_15.52288
51.47059
Index
Result Parameter
0
Overall Pass/Fail Result shows the mask test result
0:Pass, 1:Fail, –1:Not tested
1
RMS EVM pass/fail result (1.0 = fail, 0.0 = pass)
2
Peak EVM pass/fail result (1.0 = fail, 0.0 = pass)
3
Rho pass/fail result (1.0 = fail, 0.0 = pass)
4
Peak Code Domain Error pass/fail result (1.0 = fail, 0.0 = pass)
5
Frequency Error pass/fail result (1.0 = fail, 0.0 = pass)
6
RMS EVM (Average) is a floating point number (in percent) of EVM over the entire measurement area
7
Peak EVM (Peak Hold) is a floating point number (in percent) of the Peak EVM in the entire measurement
area.
8
Magnitude Error (Average) is a floating point number (in percent) of the average magnitude error over the
entire measurement area.
9
Magnitude Error (Peak Hold) is a floating point number (in percent) of the Peak magnitude error over the
258
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...