6 List Sequence Measurements
List Sequence Measurements
. . . . . .
Offset 14 (Offset O) – switching transients measurement results
3
PCL is the Power Control Level that is determined by the Transmit Power and used to determine the limits
parameter_table_23.52941
52.94118
Key Path
Meas
Initial S/W Revision
A.05.00
Help Map ID
0
GMSK Phase & Frequency (PFER) Results
The GMSK Phase & Frequency (PFER) measurement used in the GSM mode is used in the sequencer. If the
analysis step has requested this measurement then the results returned for the analysis step within the
List Sequencer measurement are listed below.
For more information on each result see the online help for the relevant radio standard measurement
mode.
The SCPI-only header values are as follows:
Meas Bitmap Header is a floating point number which appears ahead of the results for the analysis
interval; it is a decimal representation of the meas bitmap value for the measurements included in the
interval. The value returned for PFER results is 64. If more than one measurement is included in this
analysis step, then the value returned represents the combined meas bitmap value for all included
measurements, and the results are returned in ascending order of their meas bitmap values.
Number Of Results is a floating point number which appears ahead of the results of a measurement within
the analysis interval; it represents the number of results metrics returned by the measurement. The value
returned for PFER results is 17.
GMSK PFER measurement metrics return the following comma-separated scalar results:
non_
parameter_
table_
15.52288
51.47059
Index
Result Parameter
0
Overall PFER Pass/fail result. It combines all the pass fail values from the results below and OR’s them together. If
any fail this result is 1.0 = fail, 0,0 = pass
1
RMS Phase Error pass/fail result (1.0 = fail, 0,0 = pass)
2
Peak Phase Error pass/fail result (1.0 = fail, 0,0 = pass)
3
Frequency Error pass/fail result (1.0 = fail, 0,0 = pass)
4
Average RMS Phase Error
A floating point number (in degrees) of the rms phase error between the measured phase and the ideal phase over
Sequence Analyzer User's & Programmer's Reference
241
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...