6 List Sequence Measurements
Single (Single Measurement/Sweep)
Single (Single Measurement/Sweep)
Sets the analyzer for Single measurement operation. The single/continuous state is Meas Global, so the
setting will affect all the measurements. If you are Paused, pressing Single does a Resume.
See
"More Information" on page 512
Key Path
Front-panel key
Example
:INIT:CONT OFF
Notes
See
Cont
key description.
Backwards Compatibility
Notes
For Spectrum Analysis mode in ESA and PSA, the
Single
hardkey and the INITiate:IMM switched
from continuous measurement to single measurement and restarted sweeps and averages (displayed
average count reset to 1), but did not restart
Max Hold
and
Min Hold
. In the X-Series, the
Single
hardkey and the INITiate:IMM command initiate a sweep/ measurement/ average sequence/hold
sequence including
MaxHold
and
MinHold
.
For Spectrum Analysis mode in ESA and PSA, the
Single
hardkey restarted the sweep regardless of
whether or not you were in an active sweep or sweep sequence. In the X-Series,
Restart
does this
but
Single
only restarts the sweep or sweep sequence if you are in the idle state.
INIT[:IMM] in ESA & PSA Spectrum Analysis Mode does an implied ABORt. In some other PSA Modes,
INIT[:IMM] is ignored if not in the idle state. . The X-Series follows the ESA/PSA SA Mode model,
which may cause some Modes to have compatibility problems.
Initial S/W Revision
Prior to A.02.00
More Information
See
for details on the INIT:IMMediate (Restart) function.
If you are already in single sweep, the INIT:CONT OFF command has no effect.
If you are already in Single Sweep, then pressing the Single key in the middle of a sweep does not restart
the sweep or sequence. Similarly, pressing the Single key does not restart the sweep or sequence if the
sweep is not in the idle state (for example, if you are taking a very slow sweep, or the analyzer is waiting for
a trigger). Instead, it results in a message. "Already in Single, press Restart to initiate a new sweep or
sequence". Even though pressing the Single key in the middle of a sweep does not restart the sweep,
sending INIT:IMMediate does reset it.
To take one more sweep without resetting the average count, increment the average count by 1, by
pressing the step up key while Average/Hold Number is the active function, or sending the remote
command CALC:AVER:TCON UP.
512
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...