6 List Sequence Measurements
Source
MSUS (NVWFM) or a real full path representation. See example.
The segment file must reside within ARB playback memory before it can be played by the ARB player.
<reps> - Integer Type
This variable specifies the number of times a segment or sequence plays before moving on to the next
segment or sequence.
<marker> - Enum Type
NONE – This choice disables all four markers for the waveform. Disabling markers means that the
waveform sequence ignores the segments or sequence marker settings.
M1, M2, M3, M4 – these choices, either individually or a combination of them, enable the markers
for the waveform segment or sequence. Markers not specified are ignored for that segment or
sequence.
ALL – This choice enables all four markers in the waveform segment or sequence.
<waveform2> - String type.
This variable specifies the name of a second existing waveform segment. The path supports MSUS
(NVWFM) and real full path representation both. See example.
The segment file must reside within ARB playback memory before it can be played by the ARB player.
<reps> same as above, for the 2nd waveform segment.
<marker> same as above, for the 2nd waveform segment.
You can insert several waveform segments into a waveform sequence file. Just repeat inserting
waveform segments as described above.
Error Checks for Setup SCPI command:
If you do not specify a filename, or you use an unsupported MSUS (that is, not NVWFM), or have an
error in the waveform sequence file path, an error is generated.
Notes
Error Checks for Query SCPI command: (Continued)
If the specified waveform sequence file name suffix is not “.seq”, error is generated.
If you use an unsupported MSUS (that is, not NVWFM), or have an error in the waveform segment file
path, an error is generated.
If the first specified waveform file cannot be found, an error is generated.
If you nest one waveform sequence file into another waveform sequence file, an error is generated.
If the specified repetition value is larger than 65535 or smaller than 1, an error is generated.
If the specified marker type is unrecognized, an error is generated.
For Query SCPI
For the Query the parameters are:
<filename> - String type.
This variable specifies the path and name of the waveform sequence file being queried. The path
supports MSUS (NVWFM) or a real full path representation. See example.
The return value is a <string>, which includes each waveform segment file name, repetitions, and
marker type. For example:
>:SOUR:RAD:ARB:SEQ? “NVWFM:testSeq1.seq”,
<“wfmSegment1. wfm, 10, ALL, wfmSegment2.wfm, 20, M1M3”,
Error Checks for Query SCPI command:
Sequence Analyzer User's & Programmer's Reference
575
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...