3.1.5
3 - 11
Input of a temperature program requires setting of the
following items.
Start
Temp.
Dry .................... 50 to 1000
C
Ash .................... 50 to 2800
C
Atomize ............. 50 to 2800
C
Clean ................. 50 to 3000
C
(50 to 2800°C for other than Pyro
Tube C HR and D HR)
NOTE
: The maximum temperature varies with the type of
cuvette: 3000°C for Pyro Tube C HR and Pyro Tube D
HR, 2800°C for Pyro Tube HR and Tube HR, 2600 °C for
platform HR, and 2100 °C for Cup HR. Although entry
is allowed up to 2800 °C on the window, temperature
will not rise beyond the above limits due to the
relationship between cuvette shape and resistance.
End
Temp.
Set the end temperature of dry and ash stages.
You cannot enter each end temperature of atomize and
clean stages because the temperature will be
automatically set at the same temperature as for start.
Dry .................... 50 to 1000
C
Ash .................... 50 to 2800
C
Atomize ............. 50 to 2800
C
Clean ................. 50 to 3000
C
(50 to 2800°C for other than Pyro
Tube C HR and D HR)
Ramp
Time
Temperature can be raised linearly from the start
temperature to end temperature in the set time.
Input range
Dry .................... 1 to 100 s
Ash .................... 1 to 100 s
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...